In this work, the development and testing of a novel fiber-optic based label-free biosensor is presented, whose performance were verified through the detection of C-reactive protein (CRP) in serum. The device is based on a long period grating fabricated in a double cladding fiber with a W-shaped refractive index (RI) profile. As a result, the working point of the device was tuned to the mode transition region by chemical etching of the outer fiber cladding, obtaining a significant enhancement of the RI sensitivity and an excellent visibility of the grating resonances due to the mode transition in an all-silica structure. The fiber transducer was coated with a nanometric thin layer of graphene oxide in order to provide functional groups for the covalent immobilization of the biological recognition element. A very low limit of detection of about 0.15 ng/mL was obtained during the detection of CRP in serum, and a large working range (1 ng/mL – 100 μg/mL) of clinical relevance has been also achieved.

Long period grating in double cladding fiber coated with graphene oxide as high-performance optical platform for biosensing

Esposito, F;Srivastava, A;Campopiano, S;Iadicicco, A
2021-01-01

Abstract

In this work, the development and testing of a novel fiber-optic based label-free biosensor is presented, whose performance were verified through the detection of C-reactive protein (CRP) in serum. The device is based on a long period grating fabricated in a double cladding fiber with a W-shaped refractive index (RI) profile. As a result, the working point of the device was tuned to the mode transition region by chemical etching of the outer fiber cladding, obtaining a significant enhancement of the RI sensitivity and an excellent visibility of the grating resonances due to the mode transition in an all-silica structure. The fiber transducer was coated with a nanometric thin layer of graphene oxide in order to provide functional groups for the covalent immobilization of the biological recognition element. A very low limit of detection of about 0.15 ng/mL was obtained during the detection of CRP in serum, and a large working range (1 ng/mL – 100 μg/mL) of clinical relevance has been also achieved.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11367/87491
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 101
  • ???jsp.display-item.citation.isi??? 88
social impact