In this paper we report on the application of Fiber Bragg gratings (FBGs) for temperature monitoring during radiofrequency ablation (RFA) for tumor treatment. A proper configuration with several FBG sensors deployed close to the region reached by RF discharges has been developed to monitor the necrotized area. Our ex-vivo experiments on animal kidney and liver confirm that we were able to monitor the temperature with a resolution of 0.1 °C during a series of different and consecutive RF discharges with the laparoscopic bipolar radiofrequency device Habib 4x.

Temperature monitoring during thermal ablation on ex-vivo organs by Fiber Bragg gratings

Palumbo, Giovanna;Iadicicco, Agostino;Campopiano, Stefania;
2016-01-01

Abstract

In this paper we report on the application of Fiber Bragg gratings (FBGs) for temperature monitoring during radiofrequency ablation (RFA) for tumor treatment. A proper configuration with several FBG sensors deployed close to the region reached by RF discharges has been developed to monitor the necrotized area. Our ex-vivo experiments on animal kidney and liver confirm that we were able to monitor the temperature with a resolution of 0.1 °C during a series of different and consecutive RF discharges with the laparoscopic bipolar radiofrequency device Habib 4x.
2016
9781479982875
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11367/63908
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 1
social impact